Home » Products » Mask and Protective Testing Equipment » Protective Clothing Testing Equipment » GT-RA14 Dry Microbial Penetration Resistance Tester

GT-RA14 Dry Microbial Penetration Resistance Tester

Application
This dry microbial penetration resistance tester is applicable for surgical drapes, gowns and clean air suits to test the resistance of dry microbial penetration. It is composed of microbial aerosol generator, main detector, control system and other parts.

How to work?
The test is carried out on test pieces each fixed in a container. In every container except one a portion of talc contaminated with Bacillus subtilis is poured on the test piece. One container is left uncontaminated as a control. A sedimentation plate is inserted at the base of each container at a short distance below the test piece. The apparatus supporting the containers is then vibrated by a pneumatic ball vibrator. The talc that penetrates is captured on the sedimentation plate. The sedimentation plates are removed and incubated. The numbers of colonies produced are counted.

Standards
YY/T0506.2 Surgical drapes, gowns and clean air suits for patients, clinical staff and equipment - Part 2: Performance requirements and test methods
YY/T0506.5 Surgical drapes, gowns and clean air suits for patients, clinical staff and equipment - Part 5: Test methods for resistance to dry microbial penetration
ISO 22612 Clothing for protection against infectious agents - Test method for resistance to dry microbial penetration

Technical parameters
1.Mode of vibration: Pneumatic ball vibrator
2.Vibration frequency: 20800 times / minute
3.Vibration force: 650N
4.Stand dimension: 40cm×40cm×10mm
5.Stand material: Marble
6.Experimenting container: Stainless steel experimental container
7.Working positions:6
8. Equipped with one set of biosafety cabinet and one set of air compressor

8.1 The biosafety cabinet adopts four sides of negative pressure: that is, the left and right sides, rear and bottom walls of the whole device are all negative pressure air ducts, so that the working area and the external environment form a double-layer isolation of the air curtain and the box body, and the working area is surrounded by negative pressure environment to ensure that the sample does not leak.
8.2 Biosafety performance: personnel safety, microbial colony number of impact sampler ≤ 10 CFU / time, colony number of slit sampler ≤ 5 CFU / time
8.3 Airtightness of air passage: the air passage of the device is under the pressure of 500pa for 30 minutes, and the pressure drop is ≤ 10%. There is no bubble in the penetration part in soap bubble test
8.4 Air cleanliness class: Class 100, i.e. dust content (of particles ≥ 0.5um per I L air) ≤ 0.35
8.5 Noise: the height of the table top, 30cm from the front window and 38cm upward, under the condition that the ambient noise is 55dB, the arithmetic mean of the four directions is ≤ 58db (a)
8.6 Lighting: ≥ 680lux front window glass: toughened, UV resistant
8.7 UV intensity: ≥ 80uw / cm2
8.8 Differential pressure display: (30 / 2) LCD display, Ge differential pressure sensor. The number displayed on the LCD shows the pressure loss of the filter, and the error of the differential pressure sensor is ≤ 5%
8.9 Mechanical properties: the table top of the equipment shall be able to bear 25kg of pressure without obvious sinking, bending and other deformation
8.10 Electrical resistance: withstand voltage 1500V for 1 minute without breakdown
8.11 Liquid collecting tank: used to collect splashed liquid in the working area and equipped with drain valve.
8.12 Ultra high efficiency filter is suitable for filtering particles, smoke and microorganisms in the air. The filtration efficiency of imported Ultra Low Particulate Air (ULPA) is over 99.999% (adopting 0.3 μ m PAO method), which makes the cleanliness of working space reach 100 level. Equipped with a dual channel digital differential pressure display and electronic alarm system, it can dynamically display the pressure loss of exhausting and supplying air filters, the downdraft flow rate and inflow flow rate in the operation area, as well as dual detection. When the fluctuation of downflow velocity and inflow velocity exceeds 20% of its nominal value, the audible and visual alarm will prompt the downflow velocity and fluctuation to suggest replacement and maintenance of filters and fans.
8.13 When the opening height of the front window of the safety cabinet exceeds the specified height of the front window operation port, the audible and visual alarm will be given and the interlocking system will be started. When the opening height falls back to the specified opening height, the alarm interlocking system will be released automatically.

9.Power Supply: AC 220V 50Hz

Related products